New generation of BioTectum Classic Neo OLD laminar chambers

Biotectum classic neo

Second class microbiological safety laminar chamber, bio-gambling, certified for compliance with EN 12469:2002

What distinguishes us

 

COMFORT
  • The chamber design allows for the unabstention of the two-sided cleaning of the main window closing the work area
  • Front cover Cylinders
  • Front windshield and positioned at an angle relative to the worktop
  • additional armrest made of stainless steel, acid-resistant.
  • unique construction based on metal frame ensures high rigidity of the structure, which directly affects working conditions
  • Smooth edges of the slots on the Worktens
  • Dedicated, specially designed microprocessor-controlled control system
  • Number of built-in programs 3 + cleaning program BT-Clean
  • Light source white fluorescent or LED, (800 ÷ 1700) LX depending on version
  • Lighting elements located outside the Working Chamber area
  • Brightness Control – optional
  • The media connections located on the top of the chamber, allowing the chamber to be attached to walls and other laboratory equipment
  • housing - sheet steel electrostatically painted RAL 9047 colour
  • Iron gray style elements RAL 7011
CARE FOR THE NATURAL ENVIRONMENT
  • two to three high quality EC-switched motors guarantee very good laminar distribution, low power consumption and low noise emissions
  • energy consumption
  • low noise level
  • SUSPEND mode - operation on reduced parameters when the front window is closed, saving energy and components of the device
SAFETY
  • Primary filter and exhaust outlet minimum H14 class with an effectiveness of 99.995% for particles ≥ 0.3 μm
  • Additional Exhaust filter – optional
  • Filter under bench-option
  • recirculation system 30/70 % (exhaust air / recirculation air)
  • V – Shaped air curtain inlet – reduces the risk of curtain locking through the arms and eliminates the need for additional armrests
  • Interior of the working area made of stainless steel class 0H18 stainless steel (DIN 1.4301)
  • Stainless steel bowl 0H18 (DIN 1.4301) placed under the worktop
  • painted graphics (not stickers) of warning signs on the windscreen.
  • armrest for forearm mounting along the entire length of the working area with integrated gasket for precise closing of the front window
  • the internal side walls are entirely made of tempered glass
  • Solid stainless steel Laminarizer – hygienic chamber
  • All porous surfaces resulting from the process of production of polished
  • The average velocity of the airflow directed to the interior to protect the working person ≥ 0.4 m/s
  • Average airflow oriented towards the bottom to protect the product (0.25 ÷ 0.50) m/s
  • air flow velocity sensors controlling the operation of fans (automatic compensation of air flow velocity in the chamber and at the outlet in relation to changing working conditions, e.g. as a result of gradual clogging of filters)
  • diagnostic connector enabling the connection of an optional module for remote verification of the chamber operation parameters by an authorized service or the manufacturer via GSM network.
20190611 171233

intelligent management system

LCD panel

Control Panel with graphical LCD display with dedicated function keys – lighting, navigating the front window, programming basic and advanced UV lamps, fan operation function exhaust to 100% of power, function on/ Turn off the fans. The Turn & Touch selection dial makes it easy to navigate through the menu of available functions, such as Program selection, UV lamp programming, filter work counters preview, UV lamp, language change and more. Possibility to choose messages in Polish, English and German.

INFORMATION MESSAGES

 

  • Air velocity in the working area vertically downward [m/s]
  • flow velocity in the air barrier [m/s]
  • temperature in the working chamber space - active option if the chamber is equipped with a temperature sensor
  • built-in filter operating time counters, UV lamps with a function reminding of the need to replace them
  • messages for percentage consumption of UV lamps
  • graphical indicator of HEPA filter wear status
  • Percentage consumption messages for HEPA filters

ALARM MESSAGES

 

  • Fans operate at maximum efficiency
  • the air velocity of the barrier is too low outside the scope of EN 12469:2002
  • The speed of laminar flows outside the range of standard EN 12469:2002
  • UV Lamp Running Time Exceeded
  • Filter Time Exceeded
  • Operating time exceeded without service review
  • Window outside working position
  • Chamber Temperature exceeded – option – requires chamber retrofitting in temperature sensor
  • Interrupted sterilisation process
  • Inactive Chamber Mode Safe
  • Supply voltage loss has been detected

SELF-DIAGNOSIS

 

  • damage to the air mass sensor, distinguishing which sensor the damage relates to
  • Failure of functional elements related to the front window

SAFETY FUNCTION

 

  • Built-in security to prevent the UV lamp from being attached during operation of the chamber with raised front window
  • Deactivation of the buttons on the control panel at the raised front cover of the device

DECLARATION

 

  • UV lamp permanently installed, additional UV lamp – option
  • H2O2 automatic fumigation program for automatic fumigation with Phileas devices

taking care of visual impressions

Alchem_chambers/colours (3)

 

NEO_Comfort_100

COMFORT EQUIPMENT VERSION
  • rack for chamber made of closed profiles for chamber height 750 mm, for worktop 785 mm, cat.-no. no. 955-0012-0010
  • Worktop full, not split, removable
  • Stylistic elements Orange RAL 2000
  • Permanently mounted UV lamp located at the top of the working area from the front side of the
  • Device controller equipped with a program to perform the decontamination of the Chamber
  • Three electrical outlets in the work area including the two on the left and one on the right side of the Chamber
  • Port allowing the HEPA filter DOP test to be carried out
  • MODBUS RTU Communication Protocol Enabling cooperation with BMS (Building Management System)
  • RS-485 communication port, which allows several chambers to be in one bus to centrally gather information about their work and alarms
  • USB Connector for software update
  • Real Time Clock
  • Ability to adjust the display contrast

NEO_Advantage_100

ADVANTAGE VERSION OF THE EQUIPMENT

COMFORT + version

  • valve for flammable gases TOFF, cat.-no. 955-0200-0020
  • valve for connecting TOFF vacuum, cat.-no. 955-0200-0021
  • temperature sensor in the working chamber with indication of exceeding the temperature inside the chamber, cat.-no. 955-0300-0010
  • adjustment of the lighting intensity of the working area, cat.-no. 955-0300-0011

NEO_Autoprotect_100

AUTOPROTECT EQUIPMENT VERSION

COMFORT + version

  • valve for flammable gases TOFF, cat.-no. 955-0200-0020
  • valve for connecting TOFF vacuum, cat.-no. 955-0200-0021
  • temperature sensor in the working chamber with indication of exceeding the temperature inside the chamber, cat.-no. 955-0300-0010
  • adjustment of the lighting intensity of the working area, cat.-no. 955-0300-0011
  • "AutoProtect" Program – Set

    The chamber's combination with UPS and AutoProtect allows you to continue working with the chamber and gives you the time you need to properly secure the test material. AutoProtect" includes SMT1000I APC Smart-UPS 1000VA LCD 230V, cat.-no. 955-0300-0013, UPS module SMT1000I APC Smart-UPS 1000VA LCD 230V, cat.-no. 955-0300-0013

version of the implementation

Classic_Autoprotect_120BRBR DESIGN (standard) 
  • interior of the working area made of stainless steel, acid resistant 1.4301, AISI/ASTM 304, side walls made of tempered glass.
Classic_Autoprotect_120GRGR DESIGN
  • Interior of the working area made of sheet steel, electrostatically painted with antibacterial and chemically resistant paint (care coating IGP-DURA®) Side walls made of tempered glass. Table top made of stainless steel, acid resistant 1.430, AISI/ASTM 304 with IGP-DURA® coating or optionally without IGP-DURA® coating.

Product no. 955-0400-0052

Classic_Autoprotect_120MRMR DESIGN 
  • interior of the working area made of polished stainless steel acid-resistant - mirror effect, - 1.4301, AISI/ASTM 304, side walls made of tempered glass.

Product no. 955-0400-0053

2BPU 190301 5172 03